An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Electron and Probe Microscopy Part 1: SEM and TEM Designer: S.H.Kazemi IASBS University, Spring 2017

  2. AFM OM SAM 1 Å 1 nm 1 µm 1 mm 1 cm SEM STM Resolution Here are some of the techniques we will examine and a comparison of their lateral resolution capabilities.

  3. Optical Microscopy • a diffraction experiment • basic lens components • coarse/fine focus • Mon/Bin/Tri ocular schemes • working distance • adjust interpupillary distance • quantitation with reticle • image recording

  4. imaging with a simple lens 4 12/20/2019

  5. The meaning of focal length 5 12/20/2019

  6. Aberrations of electromagnetic lenses The most important ones to consider are: • Spherical aberration • Chromatic aberration • Astigmatism 6 12/20/2019

  7. spherical aberration Object plane • Arises because a simple lens is more powerful at the edge than at the centre • Is not a problem with glass lenses (can be ground to shape) • Disc of minimum confusion results instead of point focus: • Is not correctable for electromagnetic lenses 7 12/20/2019

  8. Coping with spherical aberration • Disc of minimum confusion has diameter given by: d = C α3 C = constant} • Hence reducing α gives a large reduction in d • But for optimal resolution we need large α ! • Best compromise is with α = 10-3 radians (= f/500) • Gives resolution = 0.1 nm - can not be bettered 8 12/20/2019

  9. Chromatic aberration • Light of different λbrought to different focal positions. • λ for electrons can be controlled by fixed KV and lens currents. • But λ of electrons can change by interaction with specimen ! • Rule of thumb: resolution >= (specimen thickness)/10 9 12/20/2019

  10. Astigmatism minimal confusion • Arises when the lens is more powerful in one plane than in the plane normal to it • Causes points to be imaged as short lines, which ‘flip’ through 90 degrees on passing through ‘focus’ (minimal confusion) 10 12/20/2019

  11. Astigmatism - arises from: • Inherent geometrical defects in ‘circular’ bore of lens • Inherent inhomogeneities in magnetic properties of pole piece • Build-up of contamination on bore of pole-piece and on apertures gives rise to non-conducting deposits which become charged as electron strike them Hence astigmatism is time-dependent and cannot be ‘designed out’ inevitably requires continuous correction 11 12/20/2019

  12. Astigmatism - correction • With glass optics (as in spectacles) astigmatism is corrected using an additional lens of strength and asymmetry opposed to the asymmetry of the basic (eye) lens • With electron optics, same principle employed: 1- Electrostatic Stigmator lens apposed to main lens 2- Strength & direction of its asymmetry user-variable • Only the OBJECTIVE lens needs accurate correction • Correction usually good for 1-2 hours for routine work 12 12/20/2019

  13. When we image a simple point 13 12/20/2019

  14. Q Optical Microscopy Resolution • Rayleigh equation d = 0.61 (l / N.A.) d is distance between objects that can still be distinguished, l is wavelength of light, N.A. is numerical aperture of lens = n sin(Qvertex)

  15. Definition of resolution 15 12/20/2019

  16. SEM - scanning electron microscopy 16 12/20/2019

  17. Electron emitter Scanning Electron Microscopy Electron gun 17 12/20/2019

  18. The electron gun 18 12/20/2019

  19. The electron gun A Wehnelt acts as a control grid and it also serves as a convergent electrostatic lens. The anode is biased to a high positive voltage (typically +1 to +30 kV) relative to the emitter so as to accelerate electrons from the emitter towards the anode, thus creating an electron beam that passes through the Wehnelt aperture. The Wehnelt is biased to a negative voltage (typically −200V to −300V) relative to the emitter, which is usually a tungsten filament or Lanthanum hexaboride (LaB6) hot cathode. The Wehnelt bias voltage creates a repulsive electrostatic field that condenses the cloud of primary electrons produced by the filament. 19 12/20/2019

  20. wavelength & voltage l Wavelength of electrons is determined by the accelerating voltage (V) on the filament from which they are emitted λ= 0.1*(150/V)0.5 (de Broglie, 1924) ) Therefore very high voltages (up to 100 kV) are used to produce small values of And (λ <0.005 nm) 20 12/20/2019

  21. Why high vacuum ? • Mean free path of electrons v short in air - at least 10 -5 mbar usually aimed for • Also - Tungsten filaments burn out in air - Columns must be kept dust free • Achieved by 2-fold pumping: Rotary (mechanical) pump + diffusion pump or using turbo pump 21 12/20/2019

  22. microscope column throat with valve to isolate cooling coils chimney 2 annular vents with jets of oil vapour emitted at high velocity to mechanical pump OIL electrical heater How does a diffusion pump work? 22 12/20/2019

  23. The electromagnetic lens • Works at fixed focal distance & variable focal length - Like the human eye lens, but unlike light optics • Simple rheostat can vary power of lens (varies current) • Electrons also spiral thro’ lens(effect easily observed) 23 12/20/2019

  24. Electron Gun Secondary Electron Detector Vacuum Chamber Scanning Electron Microscopy

  25. Electron microscopy • SEM - scanning electron microscopy • Tiny electron beam scanned across surface of specimen • Backscattered BSE) or secondary electrons (SE) detected • signal output to synchronized display

  26. Scanning Electron Microscopy

  27. Scanning Electron Microscopy 27 12/20/2019

  28. Scanning Electron Microscopy Electron gun Don't make x-rays - use electrons directly Wavelength: NOT  = hc/E (massless photons)  = h/(2melectronqVo) (non-relativistic)  = h/(2melectronqVo + q2Vo2/c2)1/2 relativistic  = h / (2melectronqVo + q2Vo2/c2)1/2  = 1.22639 / (Vo + 0.97845 · 10-6Vo2)1/2 (nm) & Vo(volts): 10 kV ——> 0.12 Å & 100 kV —> 0.037 Å

  29. Scanning Electron Microscopy  = h/(2melectronqVo + q2Vo2/c2)) • Effects of increasing voltage in electron gun: • Resolution increased ( decreased) • Penetration increases • Specimen charging increases (insulators) • Specimen damage increases • Image contrast decreases

  30. SEM Lens Electrons focused by Lorentz force from electromagnetic field F = qv x B effectively same as optical lenses Lenses are ring-shaped Coils generate magnetic field, electrons pass thru hollow center. Lens focal length is continuously variable & apertures control limit beam

  31. SEM- sample preparation Conducting : Little or no preparation attach to mounting stub for insertion into instrument may need to provide conductive path with Ag paint Non-conducting : Usually coat with conductive very thin layer (Au, C, Cr)

  32. Scanning Electron Microscopy Interaction volume Backscattered electrons come from whole volume (high energy) Secondary electrons come from neck only (low energy) 32 12/20/2019

  33. Scanning Electron Microscopy Interaction volume 33 12/20/2019

  34. Scanning Electron Microscopy 34 12/20/2019

  35. Elastically scattered SEM Backscattered e– Inelastically scattered Secondary electron emission SEM Relative Intensity Fraction of Incident Beam Energy Electron Re-emission 35 12/20/2019

  36. BSE vs. 2° Detection Both can be used, different information, different detection scheme. BSE Specular reflection Higher energy Encode some chemical information 2° Electrons Isotropic emission Very low energy Better structural contrast 36 12/20/2019

  37. Scanning Electron Microscopy Specimen-electron interactions What comes from specimen? Backscattered electrons Secondary electrons Fluorescent X-rays high energy compositional contrast Brightness of regions in image increases as atomic number increases (less penetration gives more backscattered electrons) low energy topographic contrast composition - EDS

  38. Schematic of a backscattered electron detector (BSD) for scanning electron microscopy (SEM) a light pipe to carry the photon signal from the scintillator inside the evacuated specimen chamber of the SEM to the photomultiplier outside the chamber The detector consists primarily of a scintillator inside a Faraday cage inside the specimen chamber of the microscope. A low positive voltage is applied to the Faraday cage to attract the relatively low energy (less than 50 eV by definition) secondary electrons. Example: Everhart-Thornley Detector (E-T detector or ET detector)

  39. Scanning Electron Microscopy Backscattered electron detector - solid state detector Electron energy up to 30-50 keV Annular around incident beam Repel secondary electrons with negative biased mesh Images are more sensitive to chemical composition (electron yield depends on atomic number) Line of sight necessary

  40. Scanning Electron Microscopy Secondary electron detector - scintillation detector • Positive bias mesh needed in front of detector to attract low energy electrons • Line of sight unnecessary 41 12/20/2019

  41. Scanning Electron Microscopy Choose correct detector- topography example Fracture surface in iron Backscattered electrons Secondary electrons 42 12/20/2019

  42. Scanning Electron Microscopy Composition - what elements present at a particular spot in specimen? Use solid state detector, and do energy scan for fluorescent X-rays 43 12/20/2019

  43. Image X-ray map Scanning Electron Microscopy Composition mapping - x-ray fluorescence Use solid state detector set for X-ray energy for a particular element in specimen 44 12/20/2019

  44. Scanning Electron Microscopy Contrast Comes from any kind of interaction with electron beam • Topography • Composition • Elements • Phases • Grain (crystal) orientation • Static Charging affects 45 12/20/2019

  45. An SEM example Trochodiscus longispinus in OM and SEM. Note improved depth of field and resolving capability of the SEM experiment.

  46. Another SEM example Changing the Y content in the Ni electrolyte bath from 1 to 5 g/L. Preferential growth directions are altered as the nucleation rates are changed by the co-depositing material.

  47. Transmission electron Microscope TEM

  48. TEM Typical Accel. volt. = 100-400 kV (some instruments - 1-3 MV) Spread broad probe across specimen - form image from transmitted electrons Diffraction data can be obtained from image area Many image types possible (BF, DF, HR, ...) - use aperture to select signal sources Main limitation on resolution: aberrations in main imaging lens Basis for magnification - strength of post- specimen lenses

  49. TEM compared to SEM

More Related

4617作文网周公解梦大全查旬紫砂壶周易成算死命免费全文下载给小公司起名姓冯取名起名大全周易 起名 公司周公解梦小女孩帮我下载解梦安阳周易专修学院怎么给公司起用英语名胖人服装店起名字名字解析周易免费命运测算免费起名大全女孩属鸡企业产品起名周末股票交易市场免费周易测名分析打评分萱吗起名杨字怎么起名苏氏取名起名大全蓉字起名搭配易奇八字算命大师马氏女孩起名字周易取名店名周易免费测算八字姓王起四个字女孩名字六爻占卜免费算命新材料 起名单名起名大全易大师起名淀粉肠小王子日销售额涨超10倍罗斯否认插足凯特王妃婚姻让美丽中国“从细节出发”清明节放假3天调休1天男子给前妻转账 现任妻子起诉要回网友建议重庆地铁不准乘客携带菜筐月嫂回应掌掴婴儿是在赶虫子重庆警方辟谣“男子杀人焚尸”国产伟哥去年销售近13亿新的一天从800个哈欠开始男孩疑遭霸凌 家长讨说法被踢出群高中生被打伤下体休学 邯郸通报男子持台球杆殴打2名女店员被抓19岁小伙救下5人后溺亡 多方发声单亲妈妈陷入热恋 14岁儿子报警两大学生合买彩票中奖一人不认账德国打算提及普京时仅用姓名山西省委原副书记商黎光被逮捕武汉大学樱花即将进入盛花期今日春分张家界的山上“长”满了韩国人?特朗普谈“凯特王妃P图照”王树国3次鞠躬告别西交大师生白宫:哈马斯三号人物被杀代拍被何赛飞拿着魔杖追着打315晚会后胖东来又人满为患了房客欠租失踪 房东直发愁倪萍分享减重40斤方法“重生之我在北大当嫡校长”槽头肉企业被曝光前生意红火手机成瘾是影响睡眠质量重要因素考生莫言也上北大硕士复试名单了妈妈回应孩子在校撞护栏坠楼网友洛杉矶偶遇贾玲呼北高速交通事故已致14人死亡西双版纳热带植物园回应蜉蝣大爆发男孩8年未见母亲被告知被遗忘张立群任西安交通大学校长恒大被罚41.75亿到底怎么缴沈阳一轿车冲入人行道致3死2伤奥运男篮美国塞尔维亚同组周杰伦一审败诉网易国标起草人:淀粉肠是低配版火腿肠外国人感慨凌晨的中国很安全男子被流浪猫绊倒 投喂者赔24万杨倩无缘巴黎奥运男子被猫抓伤后确诊“猫抓病”春分“立蛋”成功率更高?记者:伊万改变了国足氛围奥巴马现身唐宁街 黑色着装引猜测